Expected Values; Measuring Range; Limitations Of The Procedure - HemoCue Hb 301 Manuel D'utilisation

Masquer les pouces Voir aussi pour Hb 301:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Expected values

2
Children
11.0–14.0 g/dL (110–140 g/L, 6.8–8.7 mmol/L)
Woman
12.0–15.0 g/dL (120–150 g/L, 7.4–9.3 mmol/L)
Men
13.0–17.0 g/dL (130–170 g/L, 8.1–10.5 mmol/L)
Children, 2 years to teenage, gradually increase to adult values.
Due to a wide range of conditions (dietary, geographical, etc)
which affect normal values, it is recommended that each laboratory
establish its own normal range.

Measuring range

Displayed range: 0–25.6 g/dL (0–256 g/L, 0–15.9 mmol/L). The
method used in the HemoCue Hb 301 system has according to the
NCCLS document EP6 been demonstrated to be linear between 2–25
g/dL, (within 4 % difference in 2–5 g/dL and within 3 % difference in
5–25 g/dL)
.
5

Limitations of the procedure

Refer to the HemoCue Hb 301 Microcuvettes package insert
Specific performance characteristics
Within-run and Total precision
Within-run and total precision was determined according to the NCCLS
Document EP5-A
4
. The results given below in "Within-run precision"
and "Total precision" come from 3 batches of HemoCue Hb 301
Microcuvettes, 5 HemoCue Hb 301 Analyzers and 5 operators. No
recalibration was performed during the analyzing period. Commercially
available controls at 3 different levels were used. The hemoglobin
concentration was measured in duplicate twice a day, morning and
afternoon, during 20 consecutive days.
x
Control
N
g/dL
level
1
400
7.3
2
400
13.2
3
400
17.2
Accuracy
The results of the comparison studies between the HemoCue Hb 301
system and the International Council for Standardization in Haematology
method (ICSH)
are summarized in the table below. The study was
3
performed on 3 batches of HemoCue Hb 301 Microcuvettes, 4 HemoCue
Hb 301 Analyzers and by 8 operators. No recalibration was performed
during the analyzing period.
1
.
Study
N
Min
g/dL
1
700
1.6
45
Within-run
Total Precision
Precision
SD
CV
SD
CV
g/dL
g/dL
%
%
0.059
0.82
0.066
0.91
0.106
0.80
0.122
0.92
0.135
0.78
1.152
0.88
Max
Regression
Correlation
g/dL
line
coefficient (r)
23.0
Y = 0.984x
0.998
+ 0.113