10.7 Accuracy specifications

Accuracy data valid at 20°C to 25°C
LSB = least significant bit of analog to
digital converter
10.7.1 Oscilloscope
Channels 1, 2, 3, and 4:
-
±(3.5% x reading + 0.02DIV)
-
Input Range: ±200V
Frequency measurement:
-
±(0.01% + 0.02DIV)
10.7.2 Graphing Multimeter
Frequency accuracy
-
±(0.01% x reading + 3 LSB)
DC voltage amplitude
-
±(3.5% x reading + 1 LSB)
Robert Bosch GmbH
10.7.3 DVOM (DC voltage)
Range
Resolution
0.4V
0.0001V
4V
0.001V
40V
0.01V
400V
0.1V
10.7.4 DVOM (AC voltage)
Range
Resolution
0.4V
0.0001V
4V
0.001V
40V
0.01V
400V
0.1V
10.7.5 DVOM (Resistance)
Range
Resolution
400Ω
0.1Ω
4KΩ
0.001KΩ
40KΩ
0.01KΩ
400KΩ
0.1KΩ
4MΩ
0.001MΩ
40MΩ
0.01MΩ
10.8

Speaker

95 dBA at 10 cm

10.9 Trigger clamp

!
Close trigger clamp slowly. Do not snap
shut!
EN | 35
Accuracy
0.3% + 1 LSB
0.3% + 1 LSB
0.3% + 1 LSB
0.3% + 1 LSB
Accuracy
30 - 400 Hz
2% + 4 LSB
2% + 4 LSB
2% + 4 LSB
2% + 4 LSB
Accuracy
1% + 5 LSB
0.5% + 2 LSB
0.5% + 2 LSB
0.5% + 2 LSB
0.5% + 2 LSB
1% + 5 LSB