Remote Control Using A Pc - Radiometer Analytical MeterLab PHM220 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Remote Control using a PC

The PHM220 can be controlled from a PC.
Data transmission: 9600 baud, no parity, 8 data bits, 1 stop bit.
Description of messages:
Command
?
?TY
?ST
?RD t
?RD mV
?RD pH
Keyboard emulation:
}M
}C
}K
}S
}F
}P
}G
}O
}R
}L
}U
}D
60
Chapter 6
Answer
=TY PHM220 X9607 PHM220 software reference
=ST SMP
=ST pH/mV
=ST CAL
=ST GLP
=STPRG
=ST MOD
=ST INI
=RD xx.x C
=RD xxxx.x mV
=RD xx.xx pH
{M
{C
{K
{S
{F
{P
{G
{O
{R
{L
{U
{D
Comments
List of commands
Status: Sample mode
Status: pH/mV measurement
Status: Calibration mode
Status: GLP mode
Status: Edit mode
Status: Mode/date display
Status: Autotest mode
Read temperature
Read potential
Read pH
Mode
key
Edit
mode
Cal
key
Sample key
pH/mV key
Print
key
GLP
key
key

key

key
key
key

Hide quick links: