AT INTERVALS Measurements
A measurement at intervals is ideal if you wish to monitor pH or mV
continuously and have the result printed out at set intervals and
stored in the GLP table.
Edit
SAMPLE
Sample Results
AT INTERVALS
Sample Results
AUTOREAD
Stability Crit.
50
mpH/min
Stability Crit.
3
mV/min
Print Interval
30
s
Stop Print After
10
Edit SAMPLE
Using:
Select SAMPLE in the Edit display.
Press the down arrow key to continue.
Select AT INTERVALS using:
When
an extra result will be printed out and stored in the
GLP table.
The stability criterion set in this display will
determine when the electrode signal is
accepted as being stable, e.g. when the "drift"
is less than 50 mpH/min, the result will be locked
on the display.
Range avaliable: 0 to 400 mpH/min.
If a mV method has been selected, the stability
criterion is expressed in mV/min.
Range available: 0 to 50 mV/min.
In the example shown, the measurement data
will be printed out at intervals of 30 seconds.
Range available: 1 to 9999 seconds.
In the example shown, the printing out of
measurement data will automatically stop
min
after 10 minutes. You can stop printing earlier
by pressing the
Range available: 0 to 9999 minutes.
is pressed to accept the result manually,
key.
The EDIT mode