Radiometer Analytical MeterLab PHM220 Mode D'emploi page 46

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.
You can press the Print key during a measurement. The printout
will show the instrument ID and the current date and time in the
top line, followed by the pH result and the sample temperature.
4.
If AT INTERVALS has been selected in the Edit mode under
SAMPLE; refer to page 39. An example of the printout is shown
below:
*** PHM220 pH-Meter 657R00N003
Last Calibration: 16FEB95
2 Point calibration at 20.0 ˚ C: pH 7.000 (IUPAC) - 10.012 (IUPAC)
Zero pH:
7.02
10.08 pH
The measurement data will be printed out at regular time intervals:
10.08 pH
10.08 pH
**************************************************************************
5.
If the measurement is out of the calibration range ± 2 pH, a
warning message will be printed below the result. For example:
WARNING: pH out of calibration range
If the measurement is out of alarm limits, a message will be
printed below the result. For example:
ALARM: pH is higher than 10.20
ALARM: pH is lower than 0.00
If the temperature is out of alarm limits, a message will be
printed below the result. For example:
ALARM: Temperature is higher than 25˚C
ALARM: Temperature is lower than 10˚C
46
Chapter 4
10:15
Sensitivity:
100.8%
Sample No.: 1234
20.0 ˚ C
20.0 ˚ C
20.0 ˚ C
16FEB95
14:55 ***
12:50:12
12:53:12
12:56:12