Radiometer Analytical MeterLab PHM220 Mode D'emploi page 37

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Edit SAMPLE
AUTOREAD Measurements
Edit
SAMPLE
Sample Results
AUTOREAD
Sample Results
AT INTERVALS
Stability Crit.
50
mpH/min
Stability Crit.
3
mV/min
Accept Time
60 s
Using:
Select SAMPLE in the Edit display.
Press the down arrow key to continue.
Select AUTOREAD using:
The result will be locked on the display
as soon as either the stability criterion
or the accept time is reached.
Note: If the stability criterion is set to "0" the
measurement will continue until the signal is
accepted manually by pressing the
the accept time is reached (see below).
The stability criterion set in this display
will determine when the electrode signal
is accepted as being stable, e.g. when the
"drift" is less than 50 mpH/min, the result will
be locked on the display.
Range avaliable: 0 to 400 mpH/min.
If a mV method has been selected, the
stability criterion is expressed in mV/min.
Range available: 0 to 50 mV/min.
In the example shown, the signal will be
accepted and locked on the display after
60 seconds whether it is stable or not.
Enter "0" if you only want the electrode
signal to be accepted when it is stable.
Range available: 0 to 9999 seconds.
The EDIT mode
key or
37