Guidance and manufacturer's declaration – electromagnetic immunity
The IH20 is intended for use in the electromagnetic environment specified below. The customer or the user
of the IH20 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage va-
riations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level
T
EN 60601
test level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV line(s)
and neutral
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV line(s) and
neutral
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
Not applicable
70
Electromagnetic envi-
ronment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the re-
lative humidity should be
at least 30 %.
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Mains power quality
should be that of a typi-
cal commercial or hos-
pital environment. If the
user of the IH20 requires
continued operation duri-
ng power mains interrup-
tions, it is recommended
that the IH20 be powered
from an uninterruptible
power supply or a battery.
Not applicable