Calling the Info Mode
In measuring mode, press the [OK] key for approx. 3 seconds.
– If any warning or fault messages exist, the corresponding note or error codes
are displayed (see the Technical Handbook).
Press the [OK] key successively for the next display. The peak values and
the exposition values TWA and STEV will be displayed.
– If no key is pressed for 10 seconds, the device returns automatically to meas-
uring mode.
Info Off Mode
With the device switched off, press the [ + ] key. Gas name, measuring unit and
measuring range limit value will be displayed for all channels.
Pressing the [ + ] key again exits the Info Off Mode (or via timeout).
Calling the Quick Menu
In measuring mode, press the [ + ] key three times.
– If functions in the quick menu are activated using the PC software
"Dräger CC-Vision", you can select these functions using the [ + ] key. If no
functions have been activated in the quick menu, the device remains in meas-
uring mode.
Possible functions:
1. Bump test mode
2. Fresh air calibr.
3. Display and deletion of the peak values
Press the [OK] key to call the selected function.
Press the [ + ] key to cancel the active function and to switch to measuring
mode.
– If no key is pressed for 60 seconds, the device returns automatically to meas-
uring mode.
Replacing the batteries / rechargeable batteries
Do not replace the batteries / rechargeable batteries in hazardous areas.
Danger of explosion!
Batteries / rechargeable batteries are part of the Ex approval. Only the
following types may be used:
– Alkaline batteries – T4 – (not rechargeable!)
Energizer No. E91, Energizer No. EN91 (Industrial), Varta Type 4106
(power one) or Varta Type 4006 (industrial)
– NiMHy rechargeable batteries – T3 – (rechargeable)
GP 180AAHC (1800) max. 40 °C ambient temperature.
Switch off the device:
Press and hold the [OK] key and the [ + ] key simultaneously.
Loosen the screw on the power pack and remove the power pack.
For the battery holder (order no. 83 18 703):
Replace alkaline batteries or NiMHy rechargeable batteries. Ensure correct
polarity.
For the NiMH power pack T4 (order no. 83 18 704):
Completely replace the power pack.
Insert the power pack into the device and tighten the screw, the device
switches on automatically.
Do not throw used batteries into fire or try to open them by force. Danger of
explosion!
Dispose of the batteries in accordance with local regulations.