What is what
15
1
2
2
3
X-am 5000
7
6
4
0
5
1
Gas entry
2
Alarm LED
3
Buzzer
4
[OK] key
5
Power pack
6
[ + ] key
7
Display
Special symbols:
Fault
Warning
Display peak value
Show TWA
Show STEL
Bump test mode
Fresh air calibration
12
CH
4
8
%LEL
O
2
Vol%
9
CO
ppm
H
S
2
ppm
10
NH
3
ppm
12
13
14
11
8
IR interface
9
Fastening clip
10
Type plate
11
Charging contacts
12
Measured gas display
13
Measured value display
14
Special symbols
15
Tool for replacing the sensor
1-button calibration
Single gas calibration
Password required
Battery 100 % full
Battery 2/3 full
Battery 1/3 full
Battery empty
Configuration
For the individual configuration of a device with standard configuration the
device must be connected to a PC via the USB infrared cable (order no.
83 17 409) or the E-Cal system. The PC software "Dräger CC-Vision" is used to
perform the configuration.
– Changing the configuration: see the Technical Handbook.
Standard device configuration:
Dräger X-am 5000
1))
Bump test mode
1)
Fresh air calibr.
1)
Life sign
1)
Switch off
1)
LEL factor
(CH
)
(4.4 vol. % corresponds to 100 %LEL)
4
1)
Averaging time
1)
Different settings can be selected to meet customer requirements on delivery. The current setting can be
checked and changed with the software "Dräger CC-Vision".
First commissioning
Before using the device for the first time, insert the batteries supplied or a
charged NiMH power pack T4 (order no. 83 18 704), refer to chapter "Replacing
the Batteries". The Dräger X-am 5000 is ready for operation.
Operation
Switching on the device
Press and hold the [OK] key for approx. 3 seconds until the countdown
» 3 . 2 . 1 « shown in the display has expired.
– All the display segments, including the visual, audible and vibration alarms,
are activated for a short time.
– The software version is displayed.
– The device performs a self test.
– The next sensor to be calibrated/adjusted is displayed, along with the remain-
ing days to the next calibration/adjustment, e.g. » Ex %LEL CAL 20 «.
– The time until the bump test interval elapses is displayed in days, e.g.
» bt 123 «.
Quick bump test
On
On
allowed
4.4
15 minutes for STEL
8 hours for TWA