5. Emergency lowering, emergency lifting and emergency stop

Electric emergency lowering Birdie
If the hand control fails, the jib can be lowered by using the circular switch for emergency lowering. This
is located at the front of the control unit. The jib will be lowered as long as the button is pressed.
Birdie
and Birdie
Compact
TM
TM
Emergency lowering
Note! The mechanical emergency lowering system will only operate when there is a
patient in the hoist.The mechanical emergency lowering system has to be adjusted
according to the patient's weight.
and Birdie
TM
Emergency lifting
Battery capacity
Manual emergency lowering Birdie
In case of partial or total power failure, or if the battery runs down
while using the hoist, Birdie
lowering system located at the bottom of the motor.
The weight is preset for 75 kg. If the patient's weight is more or
less than 75 kg, it might be necessary to adjust the manual emergency
lowering.
Adjustment of the manual emergency lowering
Birdie
TM
If the lowering speed has to be adjusted according to the patient's
weight, then the screw in the red release handle must be adjusted.
Loosen to increase the speed and tighten to decrease the speed.
Note: Manual emergency lowering is only possible when a patient is
sitting in the hoist.
Emergency stop
In case of emergency:
Press the red button for
emergency stop.
The emergency stop
can be reset by turning
the stop button
13
Compact
TM
Jumbo Care Control Unit
Control Unit
TM
is equipped with a manual emergency
TM
Jumbo Care Control Unit
Emergency stop
Emergency lowering
Maintenance Indicator
(Refer to page 14)