9. Trouble-shooting

Only personnel having received the necessary instruction or training by Invacare
and maintenance on Birdie
Symptom
The mast seems loose
Castors are noisy
Hinge between jib and mast is
noisy during operation
The motor is not running
Motor noise, but no movement
in the piston rod
The piston rod only moves
in and not out
The control unit emits a beeping
sound during lifting, and the
jib motor stops (Birdie
Birdie
Compact).
TM
Contact your dealer if the above does not solve your problems.
Lubricate with medically clean oil, e.g. Kemitura Kem Lub KEM-WO 50, order no. 813239.
and Birdie
Compact.
TM
TM
Possible cause
The mast axis screw has become
loose
Dust or dirt in the castors
Missing lubrication
Hand control or motor
is not connected
No power on the battery
The red emergency stop
is activated
Motor is damaged
Motor is damaged
Max. load is exceeded
and
TM
Remedy
Insert the locking pin into the
mast axis screw, and tighten the
screw
Clean the castors
Lubricate the hinge
Insert the plugs fully
Charge or replace the battery
Turn the red button clockwise to
deactivate
Replace the motor
Replace the motor
Reduce the load (and the hoist
will function normally)
16
must perform service
®