HemoCue Glucose 201 RT Manuel D'utilisation page 57

Table des Matières

Publicité

Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
See NOTE 2 and NOTE 3.
Compliance level
Electromagnetic environment – guidance
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
part of the HemoCue systems, including
3 V/m
cables, than the recommended separation dis-
tance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d=1.2√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where (P) is the maximum output power
rating of the transmitter in watts (W) accor-
ding to the transmitter manufacturer and (d)
is the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey (a), should be less than the compliance
level in each frequency range (b).
Interference may occur in the vicinity of
equipment marked with the following sym-
bol.
57

Publicité

Table des Matières
loading

Table des Matières