Symptom
Explanation
E35
1. The current measurement has been incorrectly
E70–E71
Erroneous kinetics.
1. The cuvette is faulty.
2. The sample might be grossly lipemic.
ECU
1. The cuvette is empty or not filled with enough
2. No chemical reaction is identified in the blood
HHH
1. Measured value exceeds 27.8 mmol/L (500 mg/
stored and will remain as error code E35 until
the entire memory has been deleted
sample. Please note, the ECU function captures
primarily empty cuvettes, not under-filled
cuvettes.
filled cuvette.
dL) for a whole blood system and 31 mmol/L
(560 mg/dL) for a plasma equivalent system.
Action
1a. No action is normally required.
1b. If storage problems continue, the analyzer needs service.
Contact your local distributor.
1–2. Check that the system is used according to the instru ction
for use.
Fill a NEW cuvette and place it immediately in the analyzer
If the error code appears again, use a suitable laboratory refe-
rence method to analyze the specimen.
1–2. Fill a NEW cuvette and place it immediately in the analyzer.
If the ECU function appears again, contact your local distributor
or HemoCue AB. Use another HemoCue analyzer or a suitable
laboratory reference method to analyze the specimen.
1. The result is above the measuring range.
35