Symptom
Explanation
No characters on the display
1. The analyzer is not receiving power.
2. If on battery power, the batteries need to be
3. The display is out of order.
The display gives erroneous
1. The display is out of order.
characters.
2. The microprocessor is out of order.
1. The batteries need to be replaced.
The display shows "
".
2. If on AC power, the power adapter or the
The display does not switch
1. The magnet in the cuvette holder may be
from "6"and "GLU" to
2. The magnetic sensor is out of order.
three flashing dashes and
" "(ready for measuring).
replaced.
circuit board is out of order.
missing.
Action
1a. Check that the power adapter is connected to the AC power
supply.
1b. Check that the power adapter is securely connected to the
analyzer.
1c. Check that the cable is not damaged.
2.
Turn off the analyzer and replace the batteries, 4 type R6 or
AA, 1.5 V.
3.
The analyzer needs service. Contact your local distributor.
1.
The analyzer needs service. Contact your local distributor.
2.
The analyzer needs service. Contact your local distributor.
1.
Turn off the analyzer and replace the batteries, 4 type R6 or
AA, 1.5 V.
2a. Check that the power adapter is properly connected and
working.
2b. The analyzer needs service. Contact your local distributor.
1.
The analyzer needs service. Contact your local distributor.
2.
The analyzer needs service. Contact your local distributor.