Wall Mounting; Panel Mounting - Hach ORBISPHERE 51 Série Manuel D'utilisation Simplifié

Masquer les pouces Voir aussi pour ORBISPHERE 51 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
Install the device in a location and position that gives easy access to the disconnect device and its
operation.
Potential Instrument Damage. Delicate internal electronic components can be damaged by static
electricity, resulting in degraded performance or eventual failure.

4.1 Wall mounting

1. Attach the U-bracket (provided) to the wall with two screws
(not provided).
2. Tilt the instrument slightly backwards to align the bracket
pins and the insertion slots, and slide the instrument onto
the bracket as shown.
3. Insert the 2 locking screws with washers through the side
slots.
4. Adjust the instrument angle for better screen vision, and
lock both side screws.
4.2 Pipe mounting
1. Assemble the pipe mount bracket to the U-bracket, using the two screws
provided.
2. Attach this assembly to the pipe using two clamps (not provided).
3. Slide the instrument onto the bracket.
4. Insert the 2 locking screws with washers through the side slots.
5. Adjust the instrument angle for better screen vision, and lock both side
screws.

4.3 Panel mounting

Electrocution hazard. If the cable and connector for the power supply are not accessible
after installation, an accessible local disconnection means for the instrument power is
mandatory.
8 English
C A U T I O N
N O T I C E
N O T I C E
W A R N I N G