Specification
Analog voltage output
version on the
measurement board(s)
Measurement alarm
relays on the
measurement board(s)
System alarm relay on the
main board
Thermal cut off
Options
Wall and pipe mount
instrument
(H x D x W)
Panel mount instrument
(housing) (H x D x W)
1.2 K-M1100 sensor
Specification
Sample
temperature
Sample pressure
Sample flow rate
requirement
Sample types
Measurement
range
4 English
Details
0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)
Three alarm relays per measurement board; 1A-30 VAC or
0.5A-50 VDC on a resistance load.
Configurable to Normally Open [NO] or Normally Closed [NC] contacts
by changing the jumper positions.
Potential Electrocution Hazard. Connect only safety low voltage <
33 VAC RMS
One system alarm relay; 1A-30 VAC or 0.5A-50 VDC on a resistance
load.
Normally closed [NC] (NO relay also available) when instrument is
turned on.
Potential Electrocution Hazard. Connect only safety low voltage <
33 VAC RMS
Prevents ageing of sensors when exposed to high temperatures
USB host; Ethernet 10/100 Base-T
236.5 x 160 x 250 mm; Weight 4.25 kg
9.31 x 6.30 x 9.84 in.; Weight 8.82 lbs
156 (123) x 250 x 220 (214) mm; Weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in.; Weight 6.62 lbs
Details
Measurement from –5 to 50°C (23 to 122°F)
Sensor resistant to temperature from –5 to 100°C (23 to 212°F)
1 to 20 bar absolute (14.5 to 290 psia)
50 to 300 mL/min
K1100 sensor: Water only
M1100 sensor (low range): Water and beer
M1100 sensor (high range): Water, beer, wine, wort and carbonated drinks
For oxygen
measurement in
Low range sensors
High range sensors
W A R N I N G
W A R N I N G
liquid phase
0 to 2000 ppb
(dissolved. Indicative
values until 5000 ppb)
0 to 40 ppm
(dissolved)
gas phase
0 to 50 mbar or 0 to 5%
O
(at atm pressure)
2
0 to 1 bar or 0 to 100%
O
(at atm pressure)
2