Dräger X-zone 5500 Notice D'utilisation page 67

Masquer les pouces Voir aussi pour X-zone 5500:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Fault
Optical alarm signals not visible or poorly
visible.
Life signal is not functioning.
Gas detection instrument not detected.
X-am 5x00 switches off quickly; not
powered.
Device defect displayed.
Dräger X-zone 5500
Cause
Configuration or pattern incorrectly set.
Configuration is set incorrectly.
IR interface dirty.
Incompatible gas detection instrument.
Incorrect software version in the gas
detection instrument.
IR interface faulty, clip on gas detection
instrument not correctly positioned.
Holder is not correctly positioned.
Power contacts dirty or damp.
X-am 5x00 removed during operation
from X-zone 5500.
Poor contacting with X-am 5x00.
Faults, cause, remedy
Remedy
Configure the optical alarm signals using
the Dräger CC-Vision PC software (see
chapter on page 64).
Configure the life signal using the Dräger
CC-Vision PC software (see chapter on
page 64).
Clean IR interface.
Use X-am 5x00.
Have the software update performed by
DrägerService.
Close the clip on the X-am 5x00.
Re-position holder and check that it is
correctly fitted.
Clean power contacts.
Acknowledge alarm on X-zone 5500,
switch off X-zone 5500.
Clean power contacts on the X-zone and
X-am 5x00.
67

Publicité

Table des Matières
loading

Table des Matières