During Operation; Alarms (Default Settings); Concentration Pre-Alarm A1 - Dräger X-zone 5500 Notice D'utilisation

Masquer les pouces Voir aussi pour X-zone 5500:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6

During operation

During operation, the measured values for every measured
gas are displayed by the inserted X-am 5x00.
NOTICE
i
i
Refer to the instructions for use of the used X-am 5x00
for a description of the display.
The visual and acoustic life signal (green LED ring and single
tone) is issued depending on the configuration (1 -
60 seconds; default setting every 2 seconds). The life signal
can be configured using the Dräger CC-Vision PC software
(see chapter on page 64).
In the event of an alarm, the visual and the audible alarm will
be activated (see chapter 7 on page 59).
The X-zone 5500 will amplify the visual and audible alarm and
will permanently transmit the alarm information, via a wireless
connection or via cable, to further X-zone 5500 devices.
7

Alarms (default settings)

NOTICE
i
i
The
alarm
acknowledgeable) can be configured using the Dräger
CC-Vision PC software. The configuration of the X-am
5x00 critical for the correct response of the X-zone
5500.
In the case of an alarm in a group, the display of local
alarms are prioritised on all devices over third party
alarms.
The X-zone 5500 is equipped with two different alarm
generators:
Visual signal: LED ring (360°);
colours: red, green; pulsing.
Acoustic signal: Intense horn (108 dB (A) at 1 m distance/
120 dB (A) at 30 cm distance).
Triggering device:
As soon as a device detects an increased gas concentration,
this device will become the triggering device.
The triggering device forwards the alarms to all connected
devices wirelessly and/or via communication cable.
Receiving device:
All devices that receive an alarm from the triggering device
become receiving devices. The receiving devices generate a
secondary alarm. If the receiving device does not receive any
information from the triggering device, the secondary alarm on
the receiving devices is cancelled after 10 seconds.
NOTICE
i
i
The alarm-triggering device and the receiving devices
give a different visual alarm.
Dräger X-zone 5500
settings
(e.g. self-latching/
7.1

Concentration pre-alarm A1

The alarm is indicated by an intermittent
alarm message:
Display with grouped devices:
Legend:
1 Main device display
2 Local station number
3 Group size
4 Station number associated
with the measurement
display
5 Confirmation indicator
6 Alarm type A1 display/
measurement (alternating)
7 Gas name and unit
Display
with
ungrouped
devices:
Legend:
1 Local X-zone serial
number
2 Serial number of the
alarmed X-zone
3 Confirmation indicator
4 Alarm type A1 display/
measurement (alternating)
5 Gas name and unit
The pre-alarm A1 is not self-latching and stops when the
concentration has dropped below alarm threshold A1.
The triggering device at A1:
A single tone sounds and the LED ring flashes red (main
alarm).
The receiving device at A1:
A single tone sounds and the LED ring flashes red/green
(secondary alarm).
Acknowledging the pre-alarm:
Press the
key.
OK
The acoustic alarm messages will be switched off.
During operation
1
2
3
#0
2
7
H
S
ppm
2
6
A1
1
O
Vol%
2
20.9
0
ALL
5
04533112.eps
1
ARCL-0144
5
H
S
ppm
2
4
ARCM
A1
0009
ALL
3
04633112.eps
4
2
59