Faults, cause, remedy

11.1 Fault messages

Special symbol »
« and
displayed numerical
code:
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
29
30
31
32
68
Cause
X-am 5x00 with alkali supply unit.
Interrupted communication to the
X-zone 5500.
Communication error battery controller
X-zone 5500.
Main battery alarm X-am 5x00.
Battery pre-alarm X-am 5x00.
Charging current X-am 5x00 too low.
Holder (pump or X-am 5100) detected but no
pump installed.
Flow fault
Holder status change (pump or X-am 5100)
during running operation.
Check sum error program code
Check sum error operating parameter
Check sum error operating parameter
Check sum error operating parameter
Working memory test error
Faulty ADC conversion.
No contact with main device in grouping mode. Check wireless connection to group device.
Charging electronics faulty.
Battery completely discharged.
Battery main alarm of the X-zone 5500.
Battery pre-alarm of the X-zone 5500.
Device error X-am 5x00.
Alarm pattern X-am 5x00 faulty.
Incorrect software version in the gas detection
instrument.
Gas type or unit unknown.
More than 8 different gases in one group.
Modbus watchdog was not reset by the external
module.
Battery main alarm
Battery completely discharged.
Charging electronics faulty.
Communication error battery controller
X-zone 5500.
Remedy
Insert X-am 5x00 with battery power pack.
Test IR interface on X-zone 5500 and on X-am
5x00.
Contact DrägerService.
Test charging contacts on X-zone 5500 and on
X-am 5x00.
Test charging contacts on X-zone 5500 and on
X-am 5x00.
Test charging contacts on X-zone 5500 and on
X-am 5x00.
Use holder for diffusion mode.
(Operation with X-am 5100 not possible.)
Check intake hose.
Holder pump: Check intake hose and pump
filter.
Holder X-am 5100: Check gas inlet filter, cap,
hose and pump filter.
Check firm fit of the holder (pump or X-am
5100).
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Charge X-zone 5500.
Charge X-zone 5500.
Charge X-zone 5500.
Test X-am 5x00.
Contact DrägerService.
Have the software update performed by
DrägerService.
Test gas measuring devices individually to
ensure that all gases are known. Do not use
gas measuring devices with unknown gases.
Test sensor fitting within the group and replace
gas measuring devices if necessary or remove
from the group.
Check connection to the external module or
switch X-zone 5500 off.
Charge X-zone 5500.
Charge X-zone 5500.
Contact DrägerService.
Contact DrägerService.