Hach TitraLab AT1000 Série Manuel D'utilisation Simplifié page 24

Masquer les pouces Voir aussi pour TitraLab AT1000 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
Note: A license key is necessary to enable the syringe verification option on the instrument. Contact the
manufacturer or a sales representative to get a license key.
1. From the home screen, select Settings, then push Select.
2. From the settings screen, select Options, then push Select.
3. Push the left or right arrow key to select the expert mode, then push Done.
4. From the home screen, select Maintenance, then push Select.
5. From the maintenance screen, select Syringe verification, then push Select.
6. When the option is selected for the first time, enter the license key.
7. Follow the procedure supplied with the syringe verification kit. For additional information, refer to
the expanded version of this manual.
7.4 Manage the data log
To select data to view, delete or export, specify data filters
1. From the home screen, select Data log.
2. Select an option, then push Select.
Option
View data log
Export data log Exports measurement data from the system to an external device. Preview
Delete data log Removes measurement data from the system. Previews data selection before
3. Specify the data selection parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Result type
Application
Date
Operator
4. If View data log was the selected option, push View to see the selected data.
• Push the up and down arrow keys to select a line of data and push Detail to see more data
• If Electrode calibration is selected as the Result type, push the left and right arrow keys to
select the sensor if more than one sensor is installed. Push the up and down arrow keys to
select a line of data, then push the left and right arrow keys to see related graphs
• If only one Application is selected push the up and down arrow keys to select a line of data
and push Detail to see more data, or push the left and right arrow keys to see related graphs
5. If Export data log or Delete data log was the selected option, push Preview to see the selected
data, then push Export or Delete to start the procedure.
24 English
Description
Views measurement data. Select individual lines of data to view more
content.
data selection before it is exported. Make sure that an external device is
connected to the instrument (e.g., a USB key, external hard drive, etc.).
it is removed.
Description
Sets the type of result available.
Sets the available applications.
Sets the date range.
Sets the available operators.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Titralab at1102Titralab at1112Titralab at1122Titralab at1222

Table des Matières