The monitor tray option supports a monitor at the foot end of the product.
N N o o t t e e - - Make sure that the litter is flat before you attach the monitor tray option.
To attach the monitor tray option:
1. Attach the bracket back (A) and the bracket (B) (Figure 117) onto the end of the foot end foley bag hook at the foot end
of the product (C and D) (Figure 118).
2. Attach the screw knob onto the bracket back (E) (Figure 119).
3. Rotate the bracket over the socket at the foot end of the product.
4. Attach the tray support pole into the bracket (F) and into the socket at the head end or foot end of the product (Figure
120).
N N o o t t e e
• Make sure that the tray support pole is anchored as intended into the hole of the assembled bracket and the foot end
socket.
• Foot end functionality stops when you insert equipment into the sockets at the foot end of the product.
B
F F i i g g u u r r e e 1 1 1 1 7 7 – – B B r r a a c c k k e e t t a a n n d d b b r r a a c c k k e e t t b b a a c c k k
F F i i g g u u r r e e 1 1 1 1 9 9 – – A A t t t t a a c c h h i i n n g g t t h h e e s s c c r r e e w w k k n n o o b b
O O p p e e r r a a t t i i n n g g t t h h e e m m o o n n i i t t o o r r t t r r a a y y o o p p t t i i o o n n
W W A A R R N N I I N N G G
• Do not place objects that exceed 40 lb (18 kg) onto the monitor tray option.
• Do not exceed the 150 lb (68 kg) load capacity for the tray support pole.