3. Using a 3/8" drive torque wrench, torque the bolt to 75 ± 10 in-lb.
O O p p e e r r a a t t i i n n g g t t h h e e d d u u a a l l H H a a v v a a s s u u t t w w o o - - s s t t a a g g e e I I V V p p o o l l e e o o p p t t i i o o n n
C C A A U U T T I I O O N N
• Do not hang IV bags that exceed 40 lb (18 kg) onto the IV pole.
• Always make sure that the IV pole is at a low height to pass safely through door openings.
• Do not use the IV pole as a push/pull device.
To operate the IV pole option:
1. Lift and pivot the pole from the storage position.
2. Push the IV pole down until it locks into place.
3. To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its highest height position
(Figure 114).
4. Rotate the IV hangers to the desired position and hang the IV bags (B).
5. To lower the IV pole, turn the latch (C) until the telescoping portion lowers.
To store the IV pole option:
1. Lift up on the IV pole to release it from its receptacle.
2. Pivot the IV pole from its upward position.
3. Fold the IV pole down into the head end of the product.
EN
88
A
C
D
Torque item D
to 75 ± 10 in-lb
F F i i g g u u r r e e 1 1 1 1 3 3 – – D D u u a a l l H H a a v v a a s s u u t t w w o o s s t t a a g g e e I I V V p p o o l l e e
B
2141-409-007 Rev AA.0