Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment specified below.
Immunity trial
Electrostatic
± 6 kV when in contact
discharges
± 8 kV in the air
EN 61000-4-2
± 2 kV for the feed
Far transient bursts
cables
EN 61000-4-4
± 1 kV for the
input/output cables
Voltage shocks
Differential mode ± 1 kV
EN 61000-4-5
Common mode ± 2 kV
• <5% U
Dips, brief outages
• 40% U
and power voltage
• 70% U
variation
• <5% U
EN 61000-4-11
Magnetic field with
the network
3 A/m
frequency
(50/60 Hz)
Note: U
is the power voltage nominal value applied during the trial.
T
E N G L I S H
The user should make sure it is used in this environment.
CEI 60601
Compliance level
Severity level
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
N.A.
- for 10 ms
<5% U
T
- for 100 ms
<40% U
T
- for 500 ms
<70% U
T
- for 5 s
<5% U
T
3 A/m
2 0 1 0 - 0 3 - 2 6 • U s e r ' s m a n u a l • S o p r o 7 1 7 F i r s t
Electromagnetic environment
The floor should be wooden, concrete
or tile. If the floor is covered with a
synthetic
humidity should be at least 30%.
The main power supply quality should
be one of a traditional commercial or
hospital environment.
The main power supply quality should
be one of a traditional commercial or
hospital environment.
The main power supply quality should
be one of a traditional commercial or
10 ms
T
hospital environment.
100 ms
T
If the user of SOPRO device requires it
500 ms
T
to continue to operate during main
5 s
T
power
recommended SOPRO device is fed by
an inverter or a battery.
The magnetic field with the network
frequency should be at a characteristic
level of a location in a traditional
commercial or hospital environment.
Guide