10.2. IN CASE OF FAILURE

PROBLEMS
With a video monitor
No image displays on the
screen and camera LEDs are
not on.
Camera switches on but no
image displays on the screen.
An image displays on the
screen, but the quality is not
satisfactory.
An image displays, but it is not
really clear (blurry)
With a computer
No image displays on the
screen and the camera LEDs
are not on.
Camera switches on but no
image displays on the screen.
An image displays on the
screen, but the quality is not
satisfactory.
An image displays, but it is not
really clear (blurry)
20
S o p r o 7 1 7 F i r s t • U s e r ' s m a n u a l • 2 0 1 0 - 0 3 - 2 6
E N G L I S H
CAUSES
• Defective power
supply.
•connection problem.
• Defective monitor
power supply.
• Connection problem.
Monitor configuration. Check the video monitor configuration is correctly set
• Rotating ring.
• Hygienic protector.
• Defective power
supply.
•connection problem.
•Configuration
•Driver
• Connection problem.
Camera driver
configuration
• Rotating ring.
• Hygienic protector.
SOLUTIONS
1. Check the power supply is correctly connected to
the network and to the connection box.
2. Check the connecting cable is correctly connected
to the handpiece and to the connection box.
1. Check the video cable is correctly connected to the
monitor and to the connection box.
2. Check the monitor is switched on.
up (brightness, contrast, saturation, etc.)
1.Check the rotating ring is correctly positioned
(Extra oral
, Intra oral
2. Check the hygienic protector is correctly positioned
on the camera head.
1. Vérifier que l'alimentation secteur est bien
connectée au réseau et au boîtier de connexion.
2. Vérifier que le câble de liaison est bien connecté à
la pièce à main et au boîtier de connexion.
1. Check the camera is correctly set up in Sopro
Imaging (please, refer to Sopro Imaging user's
manual).
2. Check the camera is correctly detected in the
device driver (correct installation of its driver).
3. Check the USB cable coming from the DOCK is
correctly connected to the HUB.
Check the camera configuration in the Sopro Imaging
software (brightness, contrast, saturation, etc.).
Please refer to Sopro Imaging user's manual.
1. Check the rotating ring is correctly positioned
(Extra oral
, Intra oral
2. Check the hygienic protector is correctly positioned
on the camera head.
, macro ).
, macro
).