Symptom
E25
E30
E35
E50
E51
LLL
HHH
Explanation
Can be seen when starting the analyzer and
is an error that prevents further measure-
ments. The checksum of the calibration area
is not valid.
1. The instrument needs to be calibrated.
1. Fault or dirt in the optic or electronic
units.
1. The current measurement has been
incorrectly stored and will remain as
error code E35 until the entire memory
has been deleted.
1. The stepper motor causing the cuvette
holder to vibrate is out of order.
2. Wrong power adapter is used.
1. The cuvette holder is in the wrong
position when measuring.
2. Wrong power adapter is used.
1. Measuring value <5 mg/L.
1. Measuring value >150 mg/L.
Action
1. The analyzer needs service. Contact your local
HemoCue distributor.
1a. Turn off the analyzer and clean the optic unit using
the HemoCue Cleaner.
1b. The analyzer needs service. Contact your
distributor.
1a. No action is normally required.
1b. If storage problems continue, the analyzer needs
service. Contact your local HemoCue distributor.
1. The analyzer needs service. Contact your local
HemoCue distributor.
2. Check that the power adapter is recommended by
HemoCue.
1. The analyzer needs service. Contact your local
HemoCue distributor.
2. Check that the power adapter is recommended by
HemoCue.
1. The result is below the measuring range.
1. The result is above the measuring range.