Symptom
E08
E09
E10
E13
E21
E23
Explanation
The absorbance is too high
1. Light blocking item in the cuvette holder
1. This is caused by high absorbance-values
due to immunoinhibition.
This can happen during use and is an error
that prevents further measurements. A read
or write operation to the EEPROM did not
succeed.
1. The EEPROM is out of order.
This can be seen during use and is an error
that prevents further measurements. A
read or write operation to the RAM did not
succeed.
Testing a configuration area in EEPROM
returned as an erroneous checksum.
Data Error Real Time Clock.
Real Time Clock backup battery has been
drained.
Action
1a. Check that the analyzer and cuvettes are being
used according to the HemoCue Albumin 201
Operating Manual and instructions for use.
1b. The analyzer needs service. Contact your local
HemoCue distributor.
1. The albumin concentration is too high to be
quantified by the HemoCue Albumin 201 Analyzer.
1a. Turn off the analyzer and turn it on again
after 30 seconds. Take a new cuvette and repeat
the measurement.
1b. The analyzer needs service. Contact your local
HemoCue distributor.
1. The analyzer needs service. Contact your local
HemoCue distributor.
1. Turn off the analyzer and turn it on again after
30 seconds. If the problem continues, the
instrument needs service. Contact your local
HemoCue distributor.
The backup battery needs to be replaced. Contact your
distributor.