CYLINDER INSTALLATION (NOT SUPPLIED) (Figure 6)
Attach the reducers to the cylinders (see instruction manual for reducers).
For 5 lit cylinders install the shelf and attach it using the relative shelf stops.
For 10 and 14 lit cylinders position the shelf on the bottom of the unit.
The reducer pressure gauges (M) must be placed in such a way that they are visible to the
operator.
Secure the cylinders to the unit using the relative chain or spring included in the
supply.
Screw the O2 reducer (R1) tube end nut onto the flowmeter box supply nipple (I) (RIGHT
THREADING).
Screw the N
O reducer (R2) tube end nut onto the flowmeter box supply nipple (H) (LEFT
2
THREADING).
Close the rear door with the two magnets located on the top of it.
L
M
5 lt
N
O
2
Ø max ~ 140 mm
R2
L
M
R1
H
I
5 lt
O
2
Figure 6
Ø max ~ 140 mm
10 lt
10 lt
14 lt
14 lt
N
O
O
2
2
ENGLISH
11