Guidance And Manufacturer's Declaration - Kinetec Centura B&W Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
Product information

Guidance and manufacturer's declaration

Guidance and manufacturer's declaration - Electromagnetic emissions
The device Kinetec Centura B&W™ is intended for use in the electromagnetic environment specified below. The customer or the user of The device
Kinetec Centura B&W™ should ensure that it is used in such an environment.
Emissions test
Radio frequency emissions (RF)
CISPR 11
Radio frequency emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
Flicker emissions
IEC 61000-3-3
Electromedical appliances require special precautions concerning EMC. They therefore need to be installed and commissioned following the EMC
information supplied.
Electromedical appliances may be affected by mobile and portable RF communication devices.
WARNING: using cables and accessories other than those specified, except for those sold by KINETEC
lead to an increase in emissions or a decrease in The device Kinetec Centura B&W™ machine's immunity.
WARNING: The device Kinetec Centura B&W™ should not be used next to other appliances. If The device Kinetec Centura B&W™ must be used next
to other appliances, it should be under constant supervision to check that it is working normally in the given configuration.
Guidance and manufacturer's declaration - Electromagnetic immunity
The device Kinetec Centura B&W™ is intended for use in the electromagnetic environment specified below. The customer or the user of The device
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient /
±2 kV for electrical power lines
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
T
Kinetec Centura B&W™
Compliance
Group 1
Class B
Class A
Complies
Kinetec Centura B&W™ should ensure that it is used in such an environment.
IEC 60601
Test level
±6 kV - contact
±8 kV - air
±1 kV for input/output lines
±1 kV phase-to-phase
±2 kV phase-to-earth
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
3A/m
Electromagnetic environment – guidance
The device Kinetec Centura B&W™ uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The device Kinetec Centura B&W™ is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Compliance level
±6 kV - contact
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
±8 kV - air
be at least 30%.
±2 kV for electrical power
lines
Mains power quality should be that of a typical commercial or
hospital environment.
±1 kV for input/output lines
±1 kV phase-to-phase
Mains power quality should be that of a typical commercial or
hospital environment.
±2 kV phase-to-earth
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
Mains power quality should be that of a typical commercial or
(60% dip in U
)
T
hospital environment. If the user of The device Kinetec
for 5 cycles
Centura B&W™ requires continued operation during power
supply interruptions, we recommend powering The device
70% U
T
Kinetec Centura B&W™ using an uninterruptible power
(30% dip in U
)
T
supply or a battery.
for 25 cycles
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
Power frequency magnetic fields should be at levels
3A/m
characteristic of a typical location in a typical commercial or
hospital environment.
23/24
®
as replacements for internal components, may
Electromagnetic environment –
Guidance
EN

Publicité

Table des Matières
loading

Table des Matières