Product information

Guidance and manufacturer's declaration

Electromagnetic emissions
The « KINETEC Breva » is intended for use in the electromagnetic environment specified below. The customer or the user of the « KINETEC
Emissions test
Radio-Frequency emissions
CISPR 11
Radio-Frequency emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / Flicker emissions
IEC 61000-3-3
Electromagnetic immunity
The « KINETEC Breva » is intended for use in the electromagnetic environment specified below. The customer or the user of the « KINETEC
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage interruptions
CEI 61000-4-11
Voltage dips and voltage
variations on power supply input
lines
CEI 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
GB
18
Breva » should assure that it is used in such an environment.
Compliance
The « KINETEC Breva » uses RF energy only for its internal function. Therefore, its
Group 1
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The « KINETEC Breva » is suitable for use in all establishments including domestic
and those directly connected to the public low-voltage power supply network that
Class A
supplies buildings used for domestic purposes.
Complies
Breva » should assure that it is used in such an environment.
IEC 60601
Test level
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV for power supply
lines
±1 kV for input/output lines
±0,5 kV, ±1 kV differential
mode
±0,5 kV, ±1 kV, ±2 kV
common mode
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
3A/m
Electromagnetic environment – guidance
Compliance level
±2 kV, ±4 kV, ±6 kV
contact
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
±2 kV, ±4 kV, ±8 kV
relative humidity should be at least 30%.
air
±2 kV for power supply
lines
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV for input/output lines
±0,5 kV, ±1 kV differential
mode
Mains power quality should be that of a typical
commercial or hospital environment.
±0,5 kV, ±1 kV, ±2 kV
common mode
At the interruption, there is a reset of the
< 5% U
T
« KINETEC Breva ». After turning on, check up
(>95% dip in U
)
T
the operating parameters, then push START to
for 5 seconds
begin the session.
< 5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
Mains power quality should be that of a typical
(60% dip in U
)
T
commercial or hospital environment.
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
Power frequency magnetic fields should be at
3A/m
levels characteristic of a typical location in a
typical commercial or hospital environment.
Electromagnetic environment – guidance