Lower the stand until the reactor cover is sitting on top of the reactor vessel and secure it in this position by tightening
the knurled screws.
Fit the second safety switch (with cable) to the bottom (thinner) stand rod in this position and secure it so that it is
approx. 5 mm away from the top safety switch. Align the two switches with each other.
Switch on the laboratory stirrer drive at its main switch. Close the reactor cover and press the green button on the safe-
ty box. The stirrer drive will start up.
To switch off the stirrer drive, press the red button or open the reactor cover.
The IKA
®
LR-2.SI safety disconnector disconnects the laboratory reactor system from the mains power sup-
ply when the red EMERGENCY STOP button is pressed.
The safety disconnector must be actively switched on again if the power supply is interrupted. The applian-
ces cannot automatically restart.

Splinter guard LR-2.SP

CAUTION:
The guard does not provide protection against bursting!
DO NOT operate the LR-2.ST laboratory reactor at excessive
pressures even when the guard is fitted.
20
Attach the socket bracket to the top of the stand and place the multi-socket on
top
Connect the mains plugs of the stirrer drive and the disperser drive (if you are using
one) to the multi-socket on the safety disconnector.
Fix the power cables together with cable ties. When working with the IKA
module VM 600 basic it is particularly important that the power cable of the stirrer
drive is secured to prevent a restoring torque.
Connect the power supply cable of the safety disconnector to your power supply.
The limit switch trigger on the stand rod (switch distance approx. 5 mm) also breaks
the circuit if the reactor cover is open. This prevents any danger from rotating ele-
ments such as anchor stirrers or dispersing elements.
The safety disconnector must be switched back on again each time it is switched
off (use the green button). After a power failure it will also be necessary to switch
on the safety disconnector.
The Macrolon guard is mounted on the baseplate and protects the operator from
splinters and spray. It reduces the risk of injury caused by broken glass, and burns as
a result of accidentally touching the hot reactor vessel.
The splinter guard can be quickly fitted and removed again without the need for spe-
cial tools.
Place the Macrolon guard panel onto the baseplate and secure it using the two knur-
led screws on the bottom of the baseplate.
Fixing the guard
in position
®