12.4
Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity
Electromagnetic
Test level
interference
immunity tests
Conducted HF
3 V
eff
interference
150 kHz to
quantities
80 MHz
according to
IEC 61000-4-6
3 V/m
Radiated HF
80 MHz to
interference
2.7 GHz
quantities
according to
IEC 61000-4-3
Note 1: The higher frequency range applies for 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
a
The field strength of stationary transmitters such as base stations of wireless
phones and cell phones, ham radio operators, AM and FM radio and TV stations
can theoretically not always determined in advance. A study of the installation
site should be considered to determine the electromagnetic environment con-
cerning the stationary transmitter. If the measured field strength at the pro-
posed Aquilex Fluid Control System installation and operation site exceeds the
concordance levels listed above, the Aquilex Fluid Control System should be mon-
itored to document proper functionality and operation as intended. If unusual
performance characteristics are observed, additional measures may be required
such as changing orientation or the location of the Aquilex Fluid Control System.
Guidelines and Manufacturer's Statement - Electromagnetic Compatibility
Compliance
Electromagnetic environ-
ment guidelines
3 V
Portable and mobile wire-
eff
150 kHz to
less devices should not be
80 MHz
used in closer proximity to
the Aquilex Fluid Control
System (including cables/
lines) than the recom-
3 V/m
mended safety distance
80 MHz to
calculated based on the
2.7 GHz
transmitting frequency
and the applicable for-
mula. Recommended
safety distance:
d = 1.2P for 150 KHz
to 80 MHz
d = 1.2P for 80 MHz to
800 MHz
d = 2.3P for 800 MHz
to 2.7 GHz
With P as the rated output
of the transmitter in watts
[W] according to the infor-
mation provided by the
manufacturer of the trans-
mitter and d as recom-
mended safety distance in
meters [m].
The field strength of sta-
tionary transmitters for all
frequencies tested on site
a
should be lower than the
concordance level.
Interference is possible in
the proximity of devices
featuring the following
pictograph.
b
en
47