5. Position error.
Compact was not able to position the carousel. There was a difference found in the position
table and the actual measured position of the position sensor.
Check in MENU 4 - 4 [
•
If not correct select MENU 6-0 [
•
number of the pipette at the rinse station. Run MENU 7-9 [Q
6. Up or down sensor error.
Compact was not able to detect the position of the fill nozzle on the sensors.
Up sensor failure, the fill nozzle is not at the fill position.
•
Down sensor failure the fill nozzle is not at the home position.
•
Check for mechanical obstruction around the fill nozzle.
•
Fatal error, call distributor.
•
7. Rinse head up error.
The rinse head down sensor was not triggered during the movement time of the carousel.
Check the gap between the top of the rinse nozzle and the bottom of the pipette this should
•
be 1.5 to 2 mm.
Check if the sensor is correct, or readjust the sensor.
•
Fatal error, call distributor.
•
8. Measure head not home error.
Measure head is not at the home position.
Check the home sensor
•
Measure motor faulty.
•
9. Separator full error
It takes to long for the waste pump to empty the liquid separator
Check separator assembly on air leaks
•
Replace waste tubes
•
Exchange waste pump cassette
•
Clean liquid separator
•