Annexe pour l'StaRRsed Compact
Appendix - System messages Compact
The Compact generates four types of messages
System messages.
•
Test messages.
•
System time-out messages.
•
Error messages.
•
•
During normal operation the following "System messages" may occur:
1. Waiting tube
If a filled pipette is at the measuring position before the elapsed time has finished and the
•
operator is ready to fill the next pipette, the Waiting tube message will be displayed.
To continue the sample loading sequence the operator must wait until the pipette at the
•
measuring position has been measured.
2. Printer failure
When the printer is off in the display. Go to [MENU 5-4
•
Check paper feed and quantity.
•
Check printer cable connection.
•
Printer must be on-line.
•
Note: data will be stored in the buffer and can be printed afterwards
•
3. Reagents level empty message
All reagent containers have level detectors; the display shows an error that indicates which
•
reagent container(s) is (are) empty.
The expiry date of the reagent is exceeded or the container is opened longer than three
•
months.
Prepare new reagent as described in section Reagents preparation.
4. Waste bottle full message
The waste container also has a level detector. If the "Waste bottle full" message is indicated
•
on the display, the compact will stop the "rinse and fill" cycle until a new or empty container
has been installed.
[MENU 6-1 C
LEAR FATAL ERROR
•
5. Fatal separator error
The separator container has also a level detector. If the "Fatal separator error" message is
•
indicated on the display, the Compact will stop the "rinse" cycle until the separator is empty.
Page 102
]. must be carried out before the compact will start again.
]" and toggle on.
/
PRINTER ON
OFF
StaRRsed Compact Mode d'Emploi