The cause of this problem can be foam, or the waste pump is not working. The compact will
•
continue to measure and send the ESR results on time to the printer, but the rinse and fill
sequences are stopped until the error is solved.
During the start-up sequence all the positioning sensors are tested, if incorrect the instrument will
generate test messages
1. Test fill-nozzle unit.
Checks position of the fill-nozzle unit, if incorrect the unit will be re-positioned by the
•
system.
2. Test rinse-unit.
Checks position of the rinse-unit, if incorrect the unit will be re-positioned by the system.
•
3. Test measure-unit.
Checks position of the measure-unit, if incorrect the unit will be re-positioned by the system.
•
4. Test Needle-unit.
Checks position of the needle-unit, if incorrect the unit will be re-positioned by the system.
•
5. Test Diluter-unit
Check the position of the syringe, if incorrect the unit will be re-positioned by the system.
•
6. Test drive.
Checks position of the drive unit, if incorrect the unit will be re-positioned by the system.
•
7. Test Rack unit
Check the position of the Mixing motor, if incorrect the unit will be re-positioned by the
•
system
Check the rack, if incorrect the unit will be re-positioned by the system
•
During normal operation the following "System time-out" errors may occur.
These are usually fatal errors. Call distributor or your local supplier of the Compact.
1. Drive-unit.
Compact was not able to position the pipette belt within a certain time limit.
•
Check for mechanical obstructions.
•
2. Measure-unit.
Compact was not able to position the measure-unit within a certain time limit.
•
Check for mechanical obstructions.
•
3. Rinse-unit.
Compact was not able to position the rinse-unit within a certain time limit.
•