Annexe pour l'StaRRsed Compact
Check for mechanical obstructions.
•
4. Fill-nozzle unit.
Compact was not able to position the fill-nozzle unit within a certain time limit.
•
Check for mechanical obstructions.
•
5. Needle adapter.
Compact was not able to position the needle adapter within a certain time limit.
•
Check for mechanical obstructions.
•
6. Sample probe.
Compact was not able to position the sample probe within a certain time limit.
•
Check for mechanical obstructions.
•
The following "Error messages" may occur during normal operation.
1. Vacuum error.
Check if vacuum is available.
•
Check if the flow sensor is working in [MENU 4-7].
•
Fatal error, call distributor.
•
2. Vacuum stabilization error.
Compact was not able to get a stable reading during the vacuum test before aspirating the
sample.
Check for leakage on the pipette or fill nozzle.
•
Fatal error, call distributor.
•
3. Fill time error.
The fill sensor was not triggered in time.
Not enough liquid was sucked up in the pipette.
•
Insufficient sample volume.
•
No or poor vacuum, blocked needle or fill block.
•
4. Dilutor error.
Dilutor not started.
•
Can be seen in the run mode display as EDTA 001.
•
Check the value of the dilutor sensor in [MENU 4-6].
•
Check if vacuum is available.
•
Check if the flow sensor is working in [MENU 4-7].
•
Fatal error, call distributor.
•