Precaution
- Before using the Rightest™ System to test your blood glucose, please read all the instructions and
practice the tests including quality control test. ( Refer to page 36.)
- Please do the quality control test regularly to make sure the test results are accurate .
- The Rightest™ Blood Glucose Meter GM550 can only be used with the Rightest™ Blood Glucose
Test Strips GS550. The use of any other brand Strips should not be used under any circumstances.
The use of other brands of strips may give inaccurate results.
- The Rightest™ Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The
testing result is calibrated to be plasma equivalent for test with fresh capillary whole blood samples
from the fingertip, palm and forearm.
- The Rightest™ Blood Glucose Monitoring System is intended for self-testing. It should not be used
to diagnose diabetes mellitus.
- Rightest™ Blood Glucose Monitoring System has not been validated for use on neonates.
Therefore, it's not intended for use of neonates.
- Rightest™ Blood Glucose Meter hasn't been used for arterial blood testing yet, so it is still not
suitable to be used for arterial blood testing so far.
- Do the test at least 30 minutes after while entering another site with different ambient temperature.
- Pay attention to the environment protection when disposal of batteries.
- Using appliance in tropical and / or temperate climate.
4
All manuals and user guides at all-guides.com
- Not to be exposed dripping or splashing.
- The minimum blood sample size of Rightest™ Glucose Monitoring System is 1.0µl :(
1.0µL
1.5µL
Sample Size Example
Please take a minimum of 1.0µl to do the test on glucose monitoring system. Blood sample size
above 3.0µl might contaminate the test strip port and the meter.
Blood sample size below 1.0µl might cause inaccurate test result or might not start on meter
measurement. In this case, repeat the test with a new test strip.
PRECAUTOIN
- The meter will be set to mmol/L by default when sold in the Canada. When your display
shows " mg/dL " during setting or test, please contact customer service. Use of the wrong
unit of measure may lead to incorrect treatment.
- The meter and lancing device are for single patient use. Do not share them with anyone
including other family members. Do not use on multiple patients!!
User should wash hands thoroughly with soap and water after handling the meter, lancing
device, used lancets and/or tested strips.
Precaution
)
2.0µL
3.0µL
4.0µL
5