Performing a Quality Control Test
1. Take one test strip from vial and Re-cap the vial cap immediately.
2. Insert the test strip with view window, facing up, into test strip port.
3. While the blood drop symbol flashing, press and hold the main
button for over 3 seconds until the "
" symbol appears.
4. You will see blinking "
" symbol and "
screen, prompting you to apply Control Solution.
5. Shake the bottle of Control Solution well before opening the cap.
Then open the bottle and put the cap on the table.
6. Drip a drop of Control Solution on the top of the cap.
7. Gently touch sample entry of the strip with the Control Solution on the
top of the cap.
8. When you hear a beep (if volume is turned on ) please wait for the test result. The
screen will display the countdown number from 5.
38
All manuals and user guides at all-guides.com
1
2
4
5
" symbol symbol on the
6
7
9. Tightly replace the cap on the Control Solution bottle.
10. The Control Solution result appears. Compare your Quality Control Test result to
the Control Solution Range printed on the test strip vial label.
OPEN
PRECAUTION
- Your Control Solution results will not be calculated for average reading but still can be
recalled. The Control Solution Test result will be shown with "
- Our suggestion environment for Control Solution Test is 15~40 C.
- Before "
" and "
on strip because the meter is still in an internal check. If you do so, the meter will show "
and "
". Accompanied by beeps (if volume is turned on)
- The dispensing tip of the Control Solution Bottle should not touch the
sample entry area of the test strip. This may contaminate the entire bottle of
the Control Solution. Additionally, it may contaminate the meter via the test
strip port.
- Don't touch the tip of the control solution bottle. If you have touched it,
please clean up with water.
Performing a Quality Control Test
" symbol on the screen.
" appears, please don't touch the Control Solution to the sample entry
9
CLOSE
"
39