Limitations
- Store the strips in the original capped vial at temperatures between 4℃ to 30℃ ( 39℉ to 86℉ ) and
relative humidity below 90%. Do not freeze.
- Rightest™ Blood Glucose Test strips are designed for using with capillary whole blood samples.
Do not use serum or plasma samples.
- Inaccurate test results may be obtained at high altitude more than about 10000 feet ( 3048 meters )
above sea level.
- Hematocrit below 30% may cause higher results, and hematocrit above 60% may cause lower
results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Rightest™ Blood Glucose Monitoring System has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Uric acid ≧ 0.6 mmol/L
NOTE
- Acetaminophen, Ascorbic acid (Vitamin C), Dopamine, Ibuprofen, L-Dopa, Methyldopa,
Salicylic Acid, Tetracycline, Tolbutamide, Bilirubin-conjugated, Cholesterol, Creatinine,
Triglyceride, Maltose, Xylose, Galactose and Lactose ( when occurring in normal blood or
normal therapeutic concentrations ) do not significantly affect results. However, abnormally
high concentrations in blood may cause inaccurate results.
48
All manuals and user guides at all-guides.com
Measurement Technology
Measurement Calibration
Sample
Minimum Sample Volume
Measuring Range
Test Time
Memory Capacity
Power Saving
Operating Temperature
Operating Relative Humidity
Specification
Oxidase Electrochemical Sensor
Plasma
Capillary whole blood
1.0 microliters
1.1-33.3 mmol/L
5 seconds
500 blood glucose test results with date and time
Turn off automatically 2 minutes after last user action
Press the "
" button for 4 seconds.
10 ~ 40℃ ( 50 ~ 104℉ )
10 - 90%
49