Quality Control Test
Please use normal or high level control solutions tested with
System under Control Solution Mode. If the test result is within the Control Solution Range printed on
TM
the strip vial label, the Rightest Blood Glucose Monitoring System passes Quality Control Test. That
TM
means your Rightest Blood Glucose Monitoring System is working correctly.
Control Solution Range:
Test Strips
Control Solution Range mmol/L
Example of Control Solution Range
Normal:
4,6-6,3
printed on your test strip vial label.
High:
13,2-17,8
When should a Quality Control Test be performed ?
- To ensure that your meter and test strips are working together properly.
- To confirm that you are following the correct procedure.
- To prepare for your initial blood glucose test.
- To check the test strips when you open a new vial of strips.
- To check your meter after it has been dropped, damaged or exposed to liquids.
- When you suspect that your test results are inaccurate, or if your test results are not consistent with
the way you feel.
36
All manuals and user guides at all-guides.com
TM
Rightest
Blood Glucose Monitoring
NOTE
- If you want to purchase new normal or high level control solutions, please contact your
authorized Bionime representative.
The
possible reasons
your Control Solution results are out of the range :
- our Control Solution has expired or has been opened 3 months ago.
Y
- Your test strip has expired.
- You left the cap of the tst strip vial opened or the control solution off for a long time.
- You didn't perform the test proce
dure correctly.
-
Malfunction of the meter or the test strip.
If Control Solution results are out of range, your Rightest
not be working properly. Repeat the Quality Control Test. If your Control Solution results outside the
range still exist, do not use the Blood Glucose Monitoring System to test your blood glucose and
contact your Bionime Customer service.
PRECAUTION
- Each time you open a new bottle of Control Solution, write
the discard date on the label. Control Solution is good for
3 months after opening the bottle, or until the expiration
date printed on the label, whichever comes first.
Quality Control Test
TM
Blood Glucose Monitoring System may
Example of Expiration Date:
Exp. Date 2014-06
37