4.3 INSTALLATION

Important:
Do not install PIEZOTOME SOLO near to or on another device.
Do not conceal the mains power cable and the pedal cable in a cable cover or cable guide.
- Place the console on a fixed horizontal surface or a surface that is not sloped by an angle of more than 5
degrees.
- Check that the mains isolating switch (Fig. 2, item 5) is in position O (stop).
- Connect the mains power cable to the device's mains connector (Fig. 2, item 1).
- Connect the mains power cable to the electrical network socket equipped with an earthing pin.
- If necessary, connect the ground equipotential plug of your installation to the potential equalizing terminal of
your device (Fig. 2, item 6).
- Connect the control pedal to the pedal connector (Fig. 2, item 2).
- Position the pedal so that you can easily reach it with your foot.
- Plug the bracket into the bracket support (Fig. 2, item 4).
- Connect the handpiece cable to the connector (Fig. 1, item 7).
- Check that the device is sufficiently close to the working area to avoid having to stretch the cables. If
necessary, move the device closer.
- Hang the bottle or pouch containing physiological saline solution from the bracket.

4.4 FIRST START-UP

Before using the device for the first time, it is vital that you service and/or sterilize all equipment in accordance
with the procedures defined in section 11.
V – IRRIGATING SOLUTIONS
The device is not designed to administer medical substances and shall only be used with bottles or pouches
containing no more than 1 litre of physiological saline solution or sterile water.
VI – IRRIGATION LINES
The device shall be used with SATELEC irrigation lines.
VII - SETTINGS / PARAMETERS / INTERFACE
7.1 START-UP
Important:
It is recommended that you wait 4 seconds between stopping and restarting the device.
At start-up, the device performs an auto-check sequence and the screen displays programme d1 details.

7.2 PARAMETERS

The parameters are stored during the configuration of each programme and are recalled each time you select a
programme.

7.3 FLUSH / PRIME

The flush/prime function is controlled using the keyboard.
Press button
(Fig. 3, item 3) to activate the flush/prime function.
This function remains active while the button is pressed down.
Flush can be activated during use of the device. The flow rate during flush is 120ml/min.

7.4 IRRIGATION

a) Setting the irrigation flow rate
Set the irrigation flow rate by pressing the buttons
Each press adjusts the flow rate by + or – 10ml/min.
The value set is then displayed on the LCD screen (Fig. 3, item 7) in addition to a bar graph.
The flow rate is stored in the programme being used.
The flow rate can be altered during use of the device.
Piezotome Solo V6 V2
and
(Fig. 3, items 1 and 2).
8