Electromagnetic Immunity / Radiofrequency Mobile Equipment; Recommended Separation Distances - ACteon Satelec Piezotome Solo Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17

13. 3 ELECTROMAGNETIC IMMUNITY / RADIOFREQUENCY MOBILE EQUIPMENT

The device is intended for use in the electromagnetic environment specified in the table below. The user and/or
installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the device (including its
cables) at a distance less than that recommended and calculated according to the frequency and power of the
Conducted disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated
radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic
environment measurement (a), must be less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: -These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile
telephones (cellular/cordless), mobile radios, amateur radio, AM/FM radio broadcasts and TV broadcasts cannot
be determined exactly by theory. To assess the electromagnetic environment due to fixed radiofrequency
emitters, an electromagnetic environment measurement must be made. If the measured radiofrequency field
strength in the immediate environment where the product is used exceeds the compliance level specified above,
the performance of the product must be tested to verify whether it conforms to the specification. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.

13. 4 RECOMMENDED SEPARATION DISTANCES

The device is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances
are controlled.
The device user and/or installer can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile radiofrequency communications equipment emitters and the device,
according to the maximum output power of the equipment, as recommended in the table below.
Rated max. power of the emitter (W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can
be estimated using the equation applicable to the frequency of the emitter, where P is the max. power rating of
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
Piezotome Solo V6 V2
IEC
Compliance
60601
level
test level
3 V/m
150 KHz
3 V/m
to 80 MHz
3 V/m
80 MHz to
3 V/m
2.5 GHz
Separation distance in metres (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
the emitter in watts (W) according the manufacturer.
Electromagnetic environment - guidance
emitter.
Recommended separation distance:
d = 1.2
P
d = 1.2
80MHz to 800MHz.
P
d = 2.3
800MHz to 2.5GHz.
P
Where P is the maximum power rating of the emitter in
watts (W) according to the manufacturer's specifications
and d is the recommended minimum separation distance
in meters (m).
80 MHz to 800 MHz
P
0.12 m
0.38 m
1.2 m
3.8 m
12m
800 MHz to 2,5 GHz
d = 1.2
P
0.12 m
0.38 m
1.2 m
3.8 m
12m
d = 2.3
P
0.23 m
0.73 m
2.3 m
7.3 m
23m
15

Publicité

Table des Matières
loading

Table des Matières