13. 1 ELECTROMAGNETIC EMISSIONS

The device is intended for use in the electromagnetic environment specified in the table below. The user and/or
installer must ensure that the device is used in such an environment.
Emission test
RF emission - CISPR
11.
RF emission - CISPR
11.
Harmonic current
emission
IEC 61000-3-2.
Voltage fluctuation
and flicker
IEC 61000-3-3.

13. 2 ELECTROMAGNETIC IMMUNITY

The device is intended for use in the electromagnetic environment specified in the table below. The user and/or
installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2.
Electrical fast
transients
IEC 61000-4-4.
Surges
IEC 61000-4-5.
Voltage dips, short
interruptions and
voltage variations
IEC 61000-4-11.
Piezotome Solo V6 V2
Compliance
The device uses RF energy for internal operation.
Group 1
Therefore, its radiofrequency emissions are very low and are not
likely to cause any interference in nearby equipment.
Class B
The device is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the
public low voltage power supply network that supplies buildings
used for domestic purposes.
Complies
IEC 60601 test
Compliance level
level
± 6 KV contact
± 6 KV contact
± 8 KV air
± 8 KV air
± 2KV for power
± 2KV for power
supply lines
supply lines
± 1KV differential
± 1KV differential
mode ± 2KV
mode ± 2KV
common mode
common mode
<5% U
(>95% dip
T
in U
) for 0.5
<5% U
T
cycles.
in U
cycles.
40% U
(60% dip
T
in U
) for 5
40% U
T
cycles
U
T
70% U
(30% dip
70% U
T
in U
) for 25
U
T
T
cycles
<5% U
<5% U
(>95% dip
in U
T
in U
) for 250
cycles
T
cycles
Electromagnetic environment - guidance
Electromagnetic environment - guidance
Floors must be wood, concrete, cement or
tiled.
If floors are covered with synthetic material
(carpet, etc.), the relative humidity must be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment (hospital,
clinic).
Mains power quality should be that of a typical
commercial or hospital environment.
(>95% dip
T
) for 0.5
T
Mains power quality should be that of a typical
(60% dip in
T
commercial or hospital environment.
) for 5 cycles
If the use of the device requires continued
operation during power mains interruption, it is
(30% dip in
T
recommended that the product be powered
) for 25 cycles
from a separate power supply (UPS, etc.).
(>95% dip
T
) for 250
T
14