A/B. Insert column assembly into the
Inogen One® G5 concentrator. Do not
leave the column ends exposed; column
assembly should be inserted into the
Inogen One® G5 as soon as the dust caps
have been removed.
10. Push the column assembly into the
device such that the columns are
fully seated into the Inogen One® G5
concentrator. The spring loaded latch
button should fully return to the closed
position.
11. Connect the AC power supply cord to
the Inogen One® G5 and plug the power
supply AC cord into an electrical outlet.
Do Not Power on the Inogen One® G5
concentrator.
The following steps can be done by
pressing specific but ons on the device's
screen or within Inogen Connect App.
Steps from your device's display:
º
Press and hold the plus (+) and (-)
minus button for 5 seconds. The
screen will display the following
informational icon. Release button
once icon is displayed on screen.
º
Press the alert
screen will display the following
informational icons.
º
Press the power
on the Inogen One® G5, and use
normally.
Steps using Inogen Connect App.
º
If you are using Inogen Connect
App, navigate to the Advanced
screen, then to Additional
Information screen and click on
the Column Reset button.
button once and
button to turn
9A
9B
Closed and locked
10
(Actual appearance may vary,
depending on model with or
without pull handle.)
22