Notifi ation Icons
The Inogen One® G5 monitors various parameters during operation and utilizes an intelligent
alert system to indicate a malfunction of the concentrator. Mathematical algorithms and time
delays are used to reduce the probability of false alerts while still ensuring proper notific tion of
an alert condition.
If multiple alert conditions are detected, the highest priority alert will be displayed.
Note that failure to respond to the cause of an alert condition for low, medium and high priority
alerts potentially will result in discomfort or reversible minor injury only and develop within a
period of time suffici t to switch to a backup source of oxygen.
The following notific tion icons are accompanied by a single, short beep.
Description
Please Wait
Shutting Down
Hours:Minutes
Software Version:
Serial Number
Low Priority Alerts
The following low priority alerts are accompanied by a double beep and a solid yellow light.
Description
Battery Low
Attach Plug
Replace Columns
Check Battery
Oxygen Low
15
Display Icons
Condition/Action/Explanation
On/Off but on has been pressed for two
seconds. Concentrator is performing
system shut down.
!
Audible Alert button has been pressed for
HH:MM
fi e seconds.
Vx.x:SN
Display Icons
Condition/Action/Explanation
Battery power is low, with less than 10
minutes remaining. Attach external power
supply or power down and insert a fully
charged battery.
Column maintenance is required within 30 days.
Contact your equipment provider to arrange for
service.
Battery error has occurred. Check the
connection of your battery and ensure that it is
properly attached and latched on concentrator.
If battery error recurs with same battery, stop
using the battery and switch to a new battery or
remove battery and operate concentrator using
external power supply.
Concentrator is producing oxygen at a
O2
slightly low level (<82%) for a period of
10 minutes. If condition persists, contact
your equipment provider.
C