Maintenance Items
Replacement intake particle fil ers
Output Filter Replacement Kit
Inogen One® G5 columns
Note: Additional options may be available for country-specific p wer cords. To order contact
Inogen or your equipment provider.
For assistance, if needed, in setting up, using, maintaining, or to report unexpected operation or
events, contact your equipment provider or manufacturer.
Inogen One® G5
Column Change Procedure
1.
Turn off the nogen One® G5 concentrator by pressing the power button to shut
down the device.
2.
Remove the Inogen One® G5 concentrator from the carry case.
3.
Remove the battery from the Inogen One® G5 concentrator.
4.
Place the Inogen One® G5
concentrator on its side so that
the underside is visible. The metal
column assembly can be seen on
one side of the device.
5.
A. Unlock the column assembly by
pushing the latch button away from
the columns, or
B. By inserting the column tool (as
shown). See step 8 and remove top
dust cap to obtain tool.
RP-500
RP-404
RP-502
4
(Actual appearance may vary,
depending on model with or
without pull handle.)
5A
5B
20