Commissioning the test unit

4.1 Instructions on using the test
unit
To ensure that the test unit is operated safely and that
no accidents occur during operation, observe the
following points:
– The test unit must only be connected to the power
supply using the power supply unit provided.
– Before commissioning, the test unit must be
attached to the workstation using the table mount.
– It is crucial to protect the test unit from the ingress of
water and moisture. Ensure that no moisture from
splash water or leak detection agents can penetrate
into the test unit.
– The housing must only be opened by trained Dräger
personnel.
– The test unit must be sent to Dräger for calibration
once per year. Retain the supplied packaging for
future shipping.

4.2 Prerequisites for use

The PC on which the test software is to be installed
must meet the following minimum requirements:
– Processor: 1.6 GHz or higher
– RAM: 2 GB or larger
– Free disk space: >500 MB
– Operating system: Windows 10 or later
– 2 free USB ports

4.3 Prior to first use

Before the test unit can be put into operation, the
following steps must be carried out:
1. Install the test software on the PC and create the
data (see "Install the test software", page 7).
2. If applicable, fit accessories to the test unit.
3. Set up the test unit (see "Setting up the test unit",
page 8).
4.3.1 Install the test software
For installation, the user must have administrator
rights for the PC.
Instructions for use
|
Quaestor 6000/8000
1. Download the latest version of the Protector
Software from www.draeger.com/software and start
the setup file.
2. Follow the instructions in the setup program. The
required program components will be installed in the
selected folder and an entry for the program will be
created in the start menu.
3. Observe the information displayed during
installation.
4. Restart the PC and log in.
The following users are available on delivery of
the test software:
– The user "Chief" with the password "Createdata"
– The user "admin" with the password "admin"
5. If necessary, create additional user accounts for use
with the test software as well as entering details on
breathing safety products, customers and locations
(see "First steps" on the help pages of the Protector
Software (F1)).
4.3.2 Fitting the brackets for the pressure
gauge (Dräger Quaestor 8000 only)
● Fit the bracket on the right or left under the test unit
with the two screws supplied.

4.3.3 Connecting the microphone

1. Connect the supplied USB microphone to the PC.
2. Select and save the USB microphone in the test
software under Extras > Settings > Gauge
accuracy test.
3. Position the microphone near the warning device
during test.
Commissioning the test unit
|
7