Troubleshooting; Maintenance; Maintenance Intervals; Cleaning - Dräger Quaestor 6000 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

|

Troubleshooting

3. Lift the test unit by the handle at the back (2) and
pull it forwards and out from the table mount.
6
Troubleshooting
Fault
Cause
The test unit
makes a hissing
The test unit is
sound when it is
faulty
switched on.
The status LED
on the test unit
does not light up,
even though the
The test unit is
test unit is con-
faulty
nected to the
power supply
and a test has
been launched.
The LED on the
power supply
The power sup-
unit does not
ply unit is faulty
light up.
No communica-
The USB cable is
tion with the PC
faulty
software is possi-
ble (the test unit
The test unit is
is not recog-
faulty
nized).
7

Maintenance

7.1 Maintenance intervals

Work to be carried out After use Annually
Visually check the gel
face
Carry out an inspec-
1)
tion
Carry out a complete
2)
overhaul
1) Only to be carried out by authorized Dräger service techni-
cian.
10
Remedy
Contact service
personnel
Contact service
personnel
Contact service
personnel
Use a new cable
Contact service
personnel
Every
6 years
X
X
X
2) Dräger performs complete overhauls on test units in order to
ensure the availability of test units and support the workflow in
respiratory protection workshops. Test units may be subject to
different levels of wear due to the differing use of the test
units, however Dräger strictly recommends that a complete
overhaul is carried out by no later than six years.

7.2 Cleaning

WARNING
Risk of electric shock!
Penetrating moisture can cause a short circuit.
► Before cleaning the device, pull the power plug.
If required, clean the test unit with a moist soft cloth and
dry it.
For information on suitable cleaning agents and
disinfectants and their specifications, see docu-
ment 9100081 at www.draeger.com/IFU.

7.3 Replacing the gel face

1. Remove the old gel face from the test head.
2. Moisten the inside of the new gel face with water.
3. To place the new gel face onto the test head, first
position the eye with the measuring point on the test
head, and then position the other eye. Press the gel
face against the test head.
4. Ensure that the edge of the gel face is flush around
the entire test head.
5. Dräger recommends mounting a full face mask to
the test head until the moisture between the test
head and the gel face has evaporated (e.g.
overnight) and then commencing the test.
8

Transport

Only transport the test unit when it is switched off. The
test unit must not be used during transportation in a
vehicle.
For safety reasons, Dräger recommends lifting and
carrying the device with one hand on the handle and
the other hand on the high pressure port.
Instructions for use
|
Quaestor 6000/8000