Making Measurements
The analog bar graph displays the current measurement value. (In the example
above, 24 mA is represented as 125% in the 4-20 mA % scale.)
Table 2-19
% Scale measurement range
% Scale of 4-20 mA or 0-20 mA
[a] Applies to both autoranging and manual range selection.
You can change the % scale range (4-20 mA or 0-20 mA) by accessing the
multimeter's setup. See
information.
Use the % scale with a pressure transmitter, a valve positioner, or other output
actuators to measure pressure, temperature, flow, pH, or other process variables.
98
999.99%
9999.9%
DC mA measurement range
"Changing the % scale range"
Keysight U1273A/U1273AX User's Guide
[a]
30 mA or 300 mA
on page 136 for more