9.2 Replacing the turbines
3)
Fig. 6
Replacing the turbines
1 Turn off the extraction system by setting the Device On/Off switch (4) to "Off" (Fig. 5).
2 Pull out the mains plug (no Fig.).
3 Open the clamps on the sides and use the handle to lift off the upper section of the extraction system.
4 Undo the screws and remove the cover.
 9.3 Replacing the carbon brushes on page EN- 17
5 Loosen the turbine clamps and electrical connection, then remove the turbines.
6 Dispose of the turbines in accordance with local regulations (no Fig.).
 12 Disposal on page EN- 21
7 Insert the new turbines, attach the turbine clamps and establish an electrical connection (no Fig.).
8 Reverse the steps above to reassemble the system (no Fig.).
9.3 Replacing the carbon brushes
The first set of carbon brushes has a service life of approx. 800 hours with 230 V and of approximately 500 hours
with 115 V. The service life is reduced by 20% each time a carbon brush is replaced. The carbon brushes can be replaced
twice in total. After two replacements, the motor must then be replaced.
1)
Fig. 7
Replacing the carbon brushes
 Step 1 to step 4 on page EN-17
1 Loosen the clamps on the sides and remove the turbine cover.
2 Remove the carbon brush cover and replace the carbon brushes.
3 Dispose of the brushes in accordance with local regulations (no Fig.).
 12 Disposal on page EN- 21
4 Reverse the steps above to reassemble the system (no Fig.).
BAL.0731.0 • 2019-09-18
4)
2)
9 Maintenance and cleaning
5)
EN - 17