Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and this device. This device and system are NOT
Life-Supporting
This device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The
customer or user of this device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF Communications Equipment and the device as recommended below, according to the
maximum output power of the communications equipment.
Maximum Output
Power (Watts)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

ADVANCED MENU CONTROLS

With the device on, press the "Next" or "Previous" buttons on the keypad to scroll through the items on the display.
Name
PREVIOUS ITEM
NEXT ITEM
SETTING CHANGE
Please note that the values displayed are examples only.
Patient Menu Layout
The following charts show, by Mode, each Display you will scroll through using the Next or Previous buttons on the
keypad. Pressing the Down key at the Enable Menu Display will allow you to enter this submenu. Then use Next or
Previous buttons to scroll through it and press the Up button to return to main menu. Once the final Display is reached,
pressing the Next button will return you to the first Display.
NOTE-If the unlocked symbol is shown, you are able to adjust this setting. If the locked symbol is shown, this setting can
only be adjusted by your provider.
EN - 18
Recommended Separation Distances for the device (meters)
150 kHz to 80 MHz
D=(1.2)√ P
0.12
0.38
1.20
3.79
12.00
Symbol
Function
k
Pressing this button displays the previous menu option.
j
Pressing this button displays the next menu option.
h�
These buttons allow you to adjust the setting that is shown on the display.
i
80 to 800MHz
D=(1.2)√ P
0.12
0.38
1.20
3.79
12.00
800 MHz to 2.5 GHz
D=(2.3)√ P
0.23
0.73
2.30
7.27
23.00
SE-DV56-1